NANOCOLOR reagents for sample decomposition
Sample preparation for photometric analysis
Usually only dissolved compounds of a parameter are detected in water analysis. In strongly contaminated waters and industrial waste water these parameters are often bound in complexes or other structures and are therefore not directly accessible for the respective test. If it is necessary to determine the total amount of these substances, a decomposition step has to be done prior to analysis, where on most cases large amounts of organic material have to be decomposed. Within the NANOCOLOR system there are various rapid and easy methods available for conventional sample decomposition with solid reagents and kits with liquid reagents for complex matrices.
In some of the NANOCOLOR tube tests the reagents for sample preparation are already included and pre-dosed in additional test tubes next to the cuvettes. This is the perfect combination for the determination of total parameters such as total nitrogen or total chromium. Other reagents for sample preparation are available separately and are suitable for more than one parameter. After digestion the samples are then processed as described in the instructions for the respective NANOCOLOR test kit.
NANOCOLOR NanOx N – Oxidative digestion of samples containing nitrogen
NANOCOLOR NanOx N consists of an easy-to-dose solid oxidation reagent (peroxodisulfate) and a compensation reagent to eliminate interfering substances. After digestion, all inorganic and organic nitrogen compounds in the sample have been converted to nitrate and can be detected. The digestion of larger sample volumes allows a multiple determination from just one preparation.
Decomposition in heating block with NanOx N
Add decomposition reagent
Heat
Add compensation reagent
Decomposed sample
NANOCOLOR NanOx Metal – Oxidation of samples containing heavy metals
Undissolved metal ions and metal oxides are dissolved with the aid of acids and heat, metal ions are de-complexated and adsorptive or interfering substances are eliminated. Optimal recovery rates can be found in the analysis of heavy metals. NANOCOLOR NanOx Metal consists of an easy-to-dose solid oxidation reagent (peroxodisulfate) and a neutralizing reagent to adjust the pH value for the following determination of different metals. In addition to the digestion in the heating block, it is possible to digest samples in less time using a microwave.
NANOCOLOR crack set
For a more powerful and complete digestion of resistant samples we recommend to use the NANOCOLOR crack set. The included liquid reagents allow an oxidative sample preparation under acidic conditions (peroxodisulfate / sulfuric acid) and normal pressure at 100 °C in the heating block.
NANOCOLOR sludge
In Germany, the sewage sludge regulation regularizes the use of sludge as fertilizer in agriculture and in market gardens. Therein a limit for seven heavy metals is established. The determination of these parameters is possible with high accuracy using NANOCOLOR sludge (aqua regia) for digestion. A thorough training is recommended to learn the special working techniques before using the kit. Detailed instructions regarding sludge analysis can be provided free of charge.