| Standard SPE Procedure for CHROMABOND® HR-P |
|
Method:
|
Matrix
|
Application-No.:
|
|
SPE
|
-
|
304034
|
|
|
|
| Substances: |
|
aromatic compounds, phenols from water, nitroaromatics from water, pesticides from water, PAH from oil |
| Column: |
|
CHROMABOND® HR-P, 3 mL, 200 mg |
| Phase: |
|
CHROMABOND HR-P |
| REF number: |
|
730108 |
|
|
|
| Sample pretreatment: |
|
Individual sample preparation in reference to the compounds and matrix. |
|
|
|
| Conditions: |
|
Column conditioning: 3 mL methanol, then 3 mL water.
(Do not let run the column dry.)
Sample aspiration: The prepared sample is passed through the column by vacuum or pressure (max. 1000 mL sample volume)
Washing: With 3 mL methanol - water (2:98, v/v) are washed impurities
Elution: Analytes are eluted with 2 x 2 mL methanol/acetone (1:1, v/v) or dichloromethane |
|
|
|
|
|
|
|
|
|
| Author(-s): |
|
SPE department |
| Source: |
|
MACHEREY-NAGEL |
|
|
|
| Keywords: |
|
standard SPE procedure |
|
|
|
|
|
|
|
|
|
|
|
|
|
| Note: |
|
All solvent volumes are based on the amount of adsorbent (200 mg) and have to be adjusted according to the size of used cartridge. This protocol is a suggestion for standard procedures and should be optimized (solvents, volumes, adsorbent amount, ....) due to necessity of the investigated sample or compound and subsequent analysis (GC, HPLC, ...). |
| www.mn-net.com |
 |
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.
MACHEREY-NAGEL does not warrant the correctness of the applications. |
|
|
|
|