Editor
   

Standard SPE Procedure for CHROMABOND® PS-H+ (strong cation exchanger on PS/DVB)

Method:
Matrix
Application-No.:
SPE
-
304036

Substances:
Column:CHROMABOND® PS-H+, 3 ml, 200 mg
Phase:CHROMABOND PS-H+
MN catalog number:730690
Sample pretreatment:Sample is adjusted to pH 1-7 in relation to the cationic compounds
Conditions:Column conditioning: 3 ml methanol, then 3 ml water.
(Do not let run the column dry.)

Sample aspiration: The prepared sample is passed through the column by vacuum or pressure (max. 1000 ml sample volume)

Washing 1: With 3 ml AcOH / water (5:95, v/v) impurities are removed.
Washing 2 (Elution 1): With 3 ml methanol / water (20:80, v/v) neutral and acidic compounds can be removed (eluated).

Elution 2: Compounds are eluted with 2 x 2 ml methanol/acetone (1:1, v/v) + 5% NH3 or with 2 x 2 ml methanol/acetone (1:1, v/v) + 20% AcOH
Detection
Subsequent analysis:

Author(-s):SPE department
Source:MACHEREY-NAGEL

Keywords:standard SPE procedure; strong cation exchanger; SCX
Chromatogram:


Legend:All solvent volumes are based on the amount of adsorbent (200 or 500 mg) and have to be adjusted according to the size of used cartridge. This protocol is a suggestion for standard procedures and should be optimized (solvents, volumes, adsorbent amount, ....) due to necessity of the investigated sample or compound and subsequent analysis (GC, HPLC, ...).



www.mn-net.com
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.