Separation of 6 PAHs according to German Drinking Water Specification (TVO)
|
|
Method:
|
Matrix
|
Application-No.:
|
|
TLC
|
|
402400
|
|
|
|
| Substances: |
|
benzo[ghi]perylene; indeno[1,2,3-cd]pyrene; benzo[a]pyrene; benzo[b]fluoranthene; benzo[k]fluoranthene; fluoranthene |
| Plate: |
|
Glass plates Nano-SIL-PAH, 10 x 10 cm
|
| Phase: |
|
Nano-SIL-PAH |
| REF number: |
|
811050
|
|
|
|
|
|
|
|
|
|
| Conditions: |
|
Eluent: dichloromethane, -20 °C
Distance: 8 cm in 20 minutes
Application: 1 μL per 1 cm band (20 μg/mL per component) |
|
|
|
| Detection: |
|
TLC scanner, fluorescence 365 nm with 436 nm filter |
|
|
|
| Author(-s): |
|
TLC department |
| Source: |
|
MACHEREY-NAGEL |
|
|
|
| Keywords: |
|
PAHs; drinking water; TVO
|
|
|
|
|
|
|
Chromatogram:
|
|
|

|
|
|
|
| Legend: |
|
1.) Benzo[ghi]perylene
2.) Indeno[1,2,3-cd]pyrene
3.) Benzo[a]pyrene
4.) Benzo[b]fluoranthene
5.) Benzo[k]fluoranthene
6.) Fluoranthene |
| www.mn-net.com |
 |
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.