| PAH and PCB from Blood, Serum and Plasma |
|
Method:
|
Matrix
|
Application-No.:
|
|
SPE
|
blood
|
301350
|
|
|
|
| Substances: |
|
PAHs; PCBs
|
| Column: |
|
CHROMABOND® C18 ec, 1 mL, 100 mg |
| Phase: |
|
CHROMABOND® C18 ec |
| REF number: |
|
730011 |
|
|
|
| Sample pretreatment: |
|
Mix 1 mL sample with 1 mL water / n-propanol (85:15, v/v) and centrifuge for
10 min at 3000 U/min |
|
|
|
| Conditions: |
|
Column conditioning:
2 column volumes methanol, then 2 column volumes dist. water / n-propanol
(85:15, v/v)
Sample application:
Slowly force or aspirate the sample through the column
Column washing:
2 x 500 μL dist. water / n-propanol (85:15, v/v), dry in stream of air or nitrogen
Elution:
2 x 250 μL dichloromethane |
|
|
|
|
|
|
|
|
|
| Author(-s): |
|
SPE department
|
| Source: |
|
MACHEREY-NAGEL |
|
|
|
| Keywords: |
|
PAHs, PCBs, blood, serum, plasma
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| www.mn-net.com |
 |
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.