| Analysis of human hemoglobin chains |
|
Method:
|
Matrix
|
Application-No.:
|
|
HPLC
|
human blood
|
108240
|
|
|
|
| Substances: |
|
Hem; beta-globin; alpha-globin; AgammaT-globin; Ggamma-globin; AgammaI-globin |
| Column: |
|
EC 250/4 NUCLEOSIL® 300-5 C4 MPN |
| Phase: |
|
NUCLEOSIL® C4 MPN |
| REF number: |
|
720245.40
|
|
|
|
|
|
|
|
|
|
| Conditions: |
|
Eluent A: 20 % acetonitrile, 80 % water, 0.1 % trifluoroacetic acid (TFA)
Eluent B: 60 % acetonitrile, 40 % water, 0.1 % trifluoroacetic acid (TFA)
Gradient: linear 40–60 % B in 60 min
Flow rate: 1 mL/min
Injection volume: 10 μL
|
|
|
|
| Detection: |
|
UV, 220 nm |
|
|
|
| Author(-s): |
|
LC department |
| Source: |
|
MACHEREY-NAGEL |
|
|
|
| Keywords: |
|
proteins; hemoglobin; human blood
|
|
|
|
|
|
|
Chromatogram:
|
|
|

|
|
|
|
| Legend: |
|
Separation of a hemolyzate of human blood
1. Hem
2. beta-globin
3. alpha-globin
4. AgammaT-globin
5. Ggamma-globin
6. AgammaI-globin |
| www.mn-net.com |
 |
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.