Editor
   

PCB from waste oil

Method:
Matrix
Application-No.:
SPE
waste oil
301390

Substances:pcb
Column:CHROMABOND® SA / SiOH, 3 ml, 500/500 mg
Phase:Chromabond SA/SiOH
MN catalog number:730132
Sample pretreatment:
Conditions:Column conditioning: 1 ml n-hexane

Sample application:
Apply about 250 μl waste oil sample to the column and aspirate or force it into
the adsorbent with 2 x 1 ml n-hexane

Elution:
Aspirate or force another 2 x 500 μl n-hexane through the column;
collect all n-hexane fractions and if necessary adjust to a concentration
suitable for subsequent analysis by either evaporating n-hexane in a stream of
nitrogen or by dilution with n-hexane
Detection
Subsequent analysis:

Author(-s):Macherey-Nagel
Source:LC Department

Keywords:pcb; oil; waste oil
Chromatogram:


Legend:Compound recovery [%]
PCB 28 97
PCB 52 96
PCB 101 95
PCB 138 90
PCB 153 95
PCB 180 96
PCB 209 100



www.mn-net.com
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.