Method:
|
Matrix
|
Application-No.:
|
|
SPE
|
soil, sludge
|
301400
|
|
|
|
| Substances: |
|
PCB-28; PCB-52; PCB-101; PCB-138; PCB-153; PCB-180; PCB-209 |
| Column: |
|
CHROMABOND® NAN, 6 mL, 700/2000/700 mg |
| Phase: |
|
CHROMABOND® NAN |
| REF number: |
|
730149 |
|
|
|
| Sample pretreatment: |
|
Extract 2 g lyophilized sludge with 70 mL n-hexane, evaporate extract and
finally fill to 10 mL with n-hexane |
|
|
|
| Conditions: |
|
Column conditioning:
10 mL n-hexane
Sample application:
Aspirate 2 mL of the extract into the column
Elution:
Aspirate 40 mL n-hexane slowly through the column with weak vacuum, then
evaporate and fill to 5 mL with n-hexane
|
|
|
|
| Subsequent analysis: |
|
For further analysis we recommend GC
e.g., GC Appl. No. 211180 with capillary column OPTIMA® 5 MS, 50 m, 0.20 mm, 0.20 µm (REF 726210.50)
|
|
|
|
| Author(-s): |
|
SPE department
|
| Source: |
|
MACHEREY-NAGEL
|
|
|
|
| Keywords: |
|
PCBs; sludge; soil
|
|
|
|
|
|
|
Recovery rates:
|
|
PCB-28: 104 %
PCB-52: 100 %
PCB-101: 99 %
PCB-138: 98 %
PCB-153: 101 %
PCB-180: 98 %
PCB-209: 104 %
|
|
|
|
|
|
|
|
| www.mn-net.com |
 |
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.
MACHEREY-NAGEL does not warrant the correctness of the applications. |
|
|
|
|