Login
   

Separation of the PAH Standard according to EPA

Method:
Matrix
Application-No.:
HPLC

115040




Substances:
naphthalene; acenaphthylene; acenaphthene; fluorene; phenanthrene; anthracene; fluoranthene; pyrene; benz[a]anthracene; chrysene; benzo[b]fluoranthene; benzo[k]fluoranthene; benzo[a]pyrene; dibenz[ah]anthracene; benzo[ghi]perylene; indeno[1,2,3-cd]pyrene
Column:
EC 150/4 NUCLEOSIL® 100-5 C18 PAH
Phase:
NUCLEOSIL® 100-5 C18 PAH
REF number:
720923.40



Sample pretreatment:
10 μg/mL each in acetonitrile



Conditions:
Eluent A: methanol – water (80:20)
Eluent B: acetonitrile – tetrahydrofuran (93:7)
Gradient: 0 – 100 % B in 10 min, then 5 min at 100 % B
Flow rate: 1 mL/min
Pressure: 140 bar
Temperature: 20 °C



Detection:
UV, 260 nm



Author(-s):
HPLC department
Source:
MACHEREY-NAGEL



Keywords:
EPA; PAHs






Chromatogram:




Legend:
1. Naphthalene
2. Acenaphthylene
3. Acenaphthene
4. Fluorene
5. Phenantrene
6. Anthracene
7. Fluoranthene
8. Pyrene
9. Benz[a]anthracene
10. Chrysene
11. Benzo[b]fluoranthene
12. Benzo[k]fluoranthene
13. Benzo[a]pyrene
14. Dibenz[ah]anthracene
15. Benzo[ghi]perylene
16. Indeno[1,2,3-cd]pyrene
Note:

A fast separation of 16 EPA PAHs is shown with Appl. No. 123820
on NUCLEODUR® C18 PAH.




www.mn-net.com
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.
MACHEREY-NAGEL does not warrant the correctness of the applications.


 Print   
     
  
Copyright 2012 by MACHEREY-NAGEL GmbH & Co. KG   Imprint / Impressum Privacy StatementTerms Of Use