Ion exchange separation of protein standard on NUCLEOGEL® SCX

Method:
Matrix
Application-No.:
HPLC
-
108261




Substances:
myoglobin; α-chymotrypsinogen A; cytochrome c; lysozyme
Column:
VA 50/4.6 NUCLEOGEL® SCX 1000-8
Phase:
NUCLEOGEL® SCX
REF number:
719475



Sample:
protein standard



Conditions:
Eluent A: 0.02 mol/L KH2PO4, pH 6.0
Eluent B: A + 0.5 mol/L NaCl, pH 6.0
Gradient: linear, 0–100 % B in 20 min
Flow rate: 1 mL/min



Detection:
UV, 280 nm



Author(-s):
HPLC department
Source:
MACHEREY-NAGEL



Keywords:
proteins; cation ion exchange






Chromatogram:








Legend:
1  -  myoglobin
2  -  α-chymotrypsinogen A
3  -  cytochrome c
4  -  lysozyme



www.mn-net.com
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.
MACHEREY-NAGEL does not warrant the correctness of the applications.


 Print   
     
  