Method:
|
Matrix
|
Application-No.:
|
|
SPE
|
oil
|
301380
|
|
|
|
| Substances: |
|
PCB |
| Column: |
|
CHROMABOND® SiOH-H2SO4/SA, 3 ml, 500/500 mg,
CHROMABOND® SiOH, 3 ml, 500 mg |
| Phase: |
|
CHROMABOND SA/SiOH |
| MN catalog number: |
|
730125 |
|
|
|
| Sample pretreatment: |
|
Extract oil-contaminated solids with n-hexane. Homogenise other oil samples and dissolve 1.5 to 2.0 g in 50 ml n-hexane. Water which may cause turbidities can be removed with sodium sulphate. |
|
|
|
| Conditions: |
|
Determination with reference to German Standard DIN 51 527
Column conditioning:
Let 1 ml n-hexane flow through the CHROMABOND® SiOH-H2SO4/SA column.
Sample application:
Aspirate or force 500 μl sample through the CHROMABOND® SiOH-H2SO4/SA column. This phase offers better removal of interfering substances due to sulphonisation. Place CHROMABOND® SiOH-H2SO4/SA column on top of CHROMABOND® SiOH column with the aid of an adaptor (MN-cat. no. 730100) and after at least 30 seconds flush sample into the SiOH column with 2 x 1 ml n-hexane.
Elution:
Elute SiOH column with 3 x 0.5 ml n-hexane; adjust to a suitable concentration for subsequent GC analysis by evaporation of n-hexane in a stream of nitrogen or by dilution with n-hexane. |
|
|
|
|
|
|
|
|
|
| Author(-s): |
|
Macherey-Nagel |
| Source: |
|
LC Department |
|
|
|
| Keywords: |
|
PCB; oil |
|
|
|
|
|
|
|
|
|
|
|
|
|
| Recovery rates: |
|
Compound recovery [%]
PCB 28 99
PCB 52 95
PCB 101 99
PCB 138 94
PCB 153 99
PCB 180 96
PCB 209 101 |
| www.mn-net.com |
 |
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.
MACHEREY-NAGEL does not warrant the correctness of the applications. |
|
|
|
|