Login
     
  
     
Instruction leaflet:
NUCLEODEX

  
   
NUCLEODEX
Cyclodextrin-modified silica phases
for HPLC enantiomer separation
Features
Base material NUCLEOSIL® silica, particle size 5 μm, pore size 100 Å
modified cyclodextrins as chiral selectors
NUCLEODEX β-OH: β-cyclodextrin (R = H; n = 2) ∙ USP L45
Separation based on hydrogen bonds and dipole interactions between functional groups of the analyte and hydroxyl groups of the cyclodextrin
Examples for successful enantiomer separations: chlorthalidone and other compounds, which require free hydroxyl groups for enantioselective interactions
Eluent in column CH3OH – 0.1 % TEAA pH 4 (55:45)
NUCLEODEX α-PM: permethylated α-cyclodextrin (R = CH3; n = 1)
For all permethylated phases the ability to form hydrogen bonds is reduced, the hydrophobicity of the phase is increased compared to β‑OH, resulting in shorter retention times .
Examples for successful enantiomer separations: mecoprop and dichlorprop as free carboxylic acids, trans-stilbene oxide and styrene oxide
Eluent in column CH3OH – 50 mM phosphate pH 3 (70:30)
NUCLEODEX β-PM:
permethylated β-cyclodextrin (R = CH3; n = 2) ∙ USP L45
Examples for successful enantiomer separations:
mephobarbital (prominal), pesticides derivatives mecoprop-methyl et fenoxaprop-ethyl
Eluent in column CH3OH – 0.1 % TEAA pH 4 (65:35)
NUCLEODEX γ-PM:
permethylated γ-cyclodextrin (R = CH3; n = 3)
Examples for successful enantiomer separations:
steroids or other larger molecules
Eluent in column CH3OH – 0.1 % TEAA pH 4 (55:45)
NUCLEODEX phases are especially suited for the control of optical purity, but also for semipreparative separations and for the analysis of positional and cis-trans isomers.


Ordering information
    Length 200 mm   EC guard columns* CC guard columns**
EC analytical columns        
NUCLEODEX β-OH    
4 mm ID   720124.40   721171.30 721460.40
NUCLEODEX α-PM    
4 mm ID   720127.40   721469.30 721464.40
NUCLEODEX β-PM    
4 mm ID   720125.40   721176.30 721462.40
NUCLEODEX γ-PM    
4 mm ID   720752.40   721178.30 721466.40
NUCLEODEX screening kit    
consists of one CC 30/4 each with NUCLEODEX β-OH, α-PM, β-PM and γ-PM
and a CC column holder 30 mm
721920
All columns and guard column cartridges in packs of 1
* EC 4/3 guard column cartridges are used for EC columns of 4 mm ID with the Column Protection System guard column holder (REF 718966).
** CC 8/4 guard column cartridges are used for EC columns of 4 mm ID with guard column adapter EC (REF 721359).

 Print   
   
  