Digitalphotometer NANOCOLOR® 500 D |
Mircoprocessor controled filter photometer |
|
The digital photometer NANOCOLOR® 500 D is the consequent
development of our very successful photometer NANOCOLOR® 400 D. Adapted to
our customers’ requirements, today and tomorrow, it is suitable for universal use in water and
wastewater analysis. This includes municipal and industrial wastewater, drinking water, surface water,
ground water as well as cooling and boiler water. |
|
|
Save time and make your lab time easier |
 |
Fast measurements by NANOCOLOR® barcode technology
- Fully automatic, instant cuvette detection by the built-in laser scanner allows simple and fast routine analysis
- Selection of the test method and suitable wavelength, the actual measurement and storage of the measurement results are carried out automatically without having to use any other key
Measurement without cuvette slot cover
- The modern designed optics is insensitive to external light and allows straightforward measurements
Everything at a glance
- The modern, illuminated graphic display with user guidance shows all the important data at a glance.
Pre-programmed tests and photometric basic functions
- More than 100 pre-programmed tests with over 200 analysis programmes
- Straightforward call-up of all photometric basic functions, such as absorbance, factor, standard, transmission and kinetics
|
|
|
Experience precision and increase accuracy |
|
Self-explanatory user guidance
- All tests and menu items can be activated fast and easily. The photometer can be operated without any complex and time-consuming training
Manual with test instructions, presented as pictograms
- Safe test procedures without complex instructions
Highly accurate measurements by high-quality optical components
- The high-precision optic warrented accurate results
|
 |
|
|
|
Assure results and meet specifications |
 |
IQC according to ISO 9001 with NANOCONTROL NANOCHECK
- Conformance to requirements of internal quality control (IQC) and protection towards supervisors and authorities
- Fast and easy control of photometric accuracy of the NANOCOLOR® 500 D with NANOCONTROL NANOCHECK
(Cat. No. 925 701) by the user himself
Documentation of results in accordance to GLP
- Individual entries of sample number, sample location, user and dilution
Clear memory management
- GLP-conform storage of the measured results with all supplementary information such as date, time, sample number, sample location, user and dilution
- Fast and easy access to stored results and data sets
Convenient export of data
- Easy transfer of measured results and spectra to various common PC formats (e.g. MS Excel)
- Direct print-out of measured data with the NANOCOLOR® thermal printer (Cat. No. 919 16)
|
|
|
Enjoy flexibility and be prepared for the future |
|
No cuvette adapter required
- Round tubes (16 mm OD) and rectangular cuvettes (10, 20, 50 mm) can be used in the universal cuvette slot without any adapter
- Small accessories are not required
Programmable for user defined applications
- Free programming of up to 100 user defined methods
Fast photometer update - free of charge
- At any time, stay up-to-date by easy programme updating via Internet/PC
In the field and in the lab
The standard built-in, robust high-performance battery with charge regulator allows continuous operation
at 220V (110V) and up to 3000 on-site mains-independent measurements. The battery charge status is shown
in the graphic display. In the power saving mode, the photometer switches off automatically after 10,
20… 120 min. |
|
|
|
Technical data |
|
Type: |
Filter photometer with microprocessor control, self-test and auto-calibration |
Light sources: |
Tungsten lamp |
Optical system: |
Automatic filter wheel with 10 interference filters |
Wavelength: |
345 / 365 / 436 / 470 / 520 / 540 / 585 / 620 / 690 / 800 nm
plus 2 compartments for additional filters |
Wavelength accuracy: |
± 2 nm |
Wavelength calibration: |
Automatic |
Wavelength selection: |
Automatic, barcode |
Spectral bandwidth: |
10 – 12 nm |
Photometric range: |
± 3.0 E |
Measuring modes: |
More than 100 pre-programmed tests, 100 optionally programmable methods, absorbance, transmission,
factor, kinetics |
Cuvette holder: |
round tubes 14 mm ID, rectangular cuvettes 10, 20, 50 mm |
Data memory: |
500 measured data sets, GLP conform |
Display: |
graphic display, backlight, 64 x 128 pixels |
Operation: |
bar code technology, display user guidance, foil-covered push-buttons |
External light: |
Insensitive, open cuvette slot |
Interfaces: |
USB 1.1 and bidirectional serial RS 232 |
Update: |
Via Internet / PC |
Operating range: |
10-50 °C , max. 90% relative humidity (without condensation) |
Power supply: |
100 – 240V~, 50/60 Hz / 6V, 3.2 Ah via built-in battery with charge regulator and mains power
supply |
DimensionsL / B / H: |
227 x 282 x 105 mm |
Weight: |
2,4 kg |
Warranty: |
2 years |
This appliance comply with the following EC directives:
– 73/23/EEC of 19.02.1973 – Low-Voltage Directive
– 89/336/EEC of 03.05.89 (including amendment directive 92/31/EEC) – EMC Directive |
|
|
Ordering information |
|
Universal photometer NANOCOLOR® 500 D incl. software CD, manual,
dust cover, mains power supply, USB cable, data cable and calibration cuvette in stable
transport bag |
REF 919 500 |
|
 |