Login
     

Separation of Melamine and Cyanuric Acid on NUCLEODUR® HILIC
with UV and LC-MS Detection

Method:
Matrix
Application-No.:
HPLC

123070




Substances:
melamine; cyanuric acid
Column:
EC 125/2 NUCLEODUR® HILIC, 5 µm
Phase:
NUCLEODUR® HILIC
REF number:
760551.20



Sample concentration:
melamine: 10 µg/ml
cyanuric acid: 490 µg/ml



Conditions:

Eluent: acetonitrile / 10 mM ammonium formiate, pH 4 (90:10, v/v)
Flow rate: 0.2 ml/min
Temperature: 25 °C
Injection volume: 2.0 µl




Detection:

a) UV, 210 nm
b) LC-MS:

     melamine
      Mode: ESI+
      Sample temp.: 400 °C
      Needle: 3.0 kV
      Cone voltage: 50 V

     cyanuric acid
      Mode: ESI-
      Sample temp.: 400 °C
      Needle: 3.0 kV
      Cone voltage: 40 V




Author(-s):
HPLC department
Source:
MACHEREY-NAGEL, Germany 2009



Keywords:
toxic compounds; food; HILIC; LC-MS






Chromatogram:


a) UV detection





b) LC-MS detection











www.mn-net.com
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.
MACHEREY-NAGEL does not warrant the correctness of the applications.


 Print   
     
  