Redakteure
     

Separation of Paraquat and Diquat on NUCLEODUR® HILIC
with LC-MS Detection


Method:
Matrix
Application-No.:
HPLC

123060




Substances:

paraquat; diquat

Column:

EC 125/2 NUCLEODUR® HILIC, 3 µm

Phase:
NUCLEODUR® HILIC
REF number:

760531.20




Sample:
Concentration: 500 µg/ml



Conditions:

Eluent: acetonitrile / 50 mM ammonium formate, pH 3.2 (80:20 v/v)
Flow rate: 0.3 ml/min
Injection volume: 1.0 µl




Detection:

LC-MS
Mode: ESI+
Sample temp.: 450 °C
Needle: 3.0 kV
Cone voltage: 50 V

A separation with UV detection is shown by Appl. No.: 123050




Author(-s):
HPLC department
Source:
MACHEREY-NAGEL, Germany 2009



Keywords:
bipyridinium herbicides; HILIC; LC-MS






Chromatogram:









Legend:

1  -  paraquat
2  -  diquat




www.mn-net.com
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.
MACHEREY-NAGEL does not warrant the correctness of the applications.