Login
     

Separation of Paraquat and Diquat on NUCLEODUR® HILIC

Method:
Matrix
Application-No.:
HPLC

123050




Substances:
paraquat; diquat
Column:

EC 125/2 NUCLEODUR® HILIC, 3 µm

Phase:

NUCLEODUR® HILIC

REF number:

760531.20




Sample:
Concentration: 0.5 mg/ml



Conditions:

Eluent: acetonitrile / 50 mM ammonium formate, pH 3.2 (80:20 v/v)
Flow rate: 0.3 ml/min
Temperature: 45 °C
Injection volume: 1.0 µl




Detection:

UV, 254 nm and 308 nm

A separation with LC-MS detection is shown by Appl. No.: 123060




Author(-s):
HPLC department
Source:

MACHEREY-NAGEL, Germany 2009

acc. to EPA 549.1 and 549.2




Keywords:
bipyridinium herbicides; HILIC; EPA 549.1; EPA 549.2






Chromatogram:









Legend:
1  -  paraquat
2  -  diquat



www.mn-net.com
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.