Separation of Chlormequat and Mepiquat on NUCLEODUR® HILIC
with LC-MS Detection
|
|
Method:
|
Matrix
|
Application-No.:
|
|
HPLC
|
|
123040
|
|
|
|
| Substances: |
|
chlormequat; mepiquat |
| Column: |
|
EC 125/2 NUCLEODUR® HILIC, 3 µm
|
| Phase: |
|
NUCLEODUR® HILIC |
| REF number: |
|
760531.20 |
|
|
|
| Sample: |
|
Concentration: 500 ng/µl |
|
|
|
| Conditions: |
|
Eluent: acetonitrile / 25 mM ammonium acetate, pH 6.82 (80:20, v/v)
Flow rate: 0.3 ml/min
Temperature: 25 °C
Injection volume: 1 µl
|
|
|
|
| Detection: |
|
LC-MS
Mode: ESI+
Sample temp.: 450 °C
Needle: 3.0 kV
Cone voltage: 50 V |
|
|
|
| Author(-s): |
|
HPLC department |
| Source: |
|
MACHEREY-NAGEL, Germany 2009 |
|
|
|
| Keywords: |
|
pesticides; HILIC; LC-MS |
|
|
|
|
|
|
Chromatogram:
|
|
|

|
|
|
|
| Legend: |
|
1 - Chlormequat
2 - Mepiquat |
| www.mn-net.com |
 |
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.
MACHEREY-NAGEL does not warrant the correctness of the applications. |
|
|
|
|