Login
     

Separation of Creatinine and Creatine on NUCLEODUR® HILIC 
with LC-MS Detection

Method:
Matrix
Application-No.:
HPLC

123000




Substances:

creatinine; creatine

Column:

EC 125/2 NUCLEODUR® HILIC, 3 µm

Phase:
NUCLEODUR® HILIC
REF number:
760531.20



Sample:
Concentration: 30 ng/µL



Conditions:

Eluent: acetonitrile / 10 mM ammonium acetate, pH 4 (70:30, v/v)
Flow rate: 0.2 mL/min
Temperature: 25 °C
Injection volume: 5 µL




Detection:

LC-MS
Mode: ESI+
Sample temp.: 450 °C
Needle: 3.0 kV
Cone voltage: 50 V

A separation with UV detection is shown by Appl. No.: 122990




Author(-s):
HPLC department
Source:
MACHEREY-NAGEL, Germany 2009



Keywords:

muscle metabolism; HILIC; LC-MS







Chromatogram:









Legend:
1  -  creatinine
2  -  creatine



www.mn-net.com
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.