Login

Separation of nucleotides on NUCLEODUR® HILIC

Method:
Matrix
Application-No.:
HPLC

122920




Substances:
adenosine; cAMP; AMP; ADP; ATP
Column:

EC 125/4 NUCLEODUR® HILIC, 3 µm

Phase:
NUCLEODUR® HILIC
REF number:
760531.40









Conditions:

Eluent: acetonitrile – 100 mmol/L ammonium acetate, pH 5.3 (70:30, v/v)
Flow rate: 1.3 mL/min
Temperature: 25 °C
Injection volume: 0.5 µL




Detection:
UV, 259 nm



Author(-s):
HPLC department
Source:
MACHEREY-NAGEL, Germany 2009



Keywords:
nucleic acid bases; nucleotides; HILIC






Chromatogram:














www.mn-net.com
Concerning questions to literature applications please contact the cited author/s. Chemicals mentioned in applications are not manufactured or sold by MACHEREY-NAGEL. Applications mentioned in MACHEREY-NAGEL literature or databases are provided for informational purposes only. MACHEREY-NAGEL does not warrant that all applications have been tested in MACHEREY-NAGEL laboratories using MACHEREY-NAGEL products.
MACHEREY-NAGEL does not warrant the correctness of the applications.