Criteria for an efficient and economical preparative HPLC

 

Preparative HPLC is one of the most important methods for the efficient clean-up of chemical and biological substances from synthesis or extraction mixtures. It is used for separation of few milligrammes of valuable end or intermediate products on the laboratory scale up to gramme and kilogramme amounts in pilot plants or production facilities. Often, similar purities cannot be achieved with other clean-up methods, like e.g. recrystallization.

Several criteria are important for selection of suitable columns or adsorbents in up-scaling from analytical to economical preparative HPLC.


  Requirements for preparative separations
  Whereas the target of analytical HPLC is a separation as complete as possible of single components of a mixture with subsequent quantification of the analytes, the demands on preparative separation are the isolation of the desired product with a defined purity and a maximum yield by an economical mode of operation. Thus, the demands in terms of purity and yield of the products, as well as the throughput of samples should be considered for the selection of a suitable preparative HPLC system.

 
High purity of the product can be achieved by high resolution between the product and the neighbouring peaks. The resolution of two peaks is determined by the chromatographic system of mobile and stationary phase. A smaller particle size of the stationary phase (e.g. 5 or 10 µm) combined with the adjustment of further system parameters facilitates higher resolution and thus higher purity.

Defined yield is also important for preparative clean-up. In particular every step in multiple-step syntheses should have a high yield for the next reaction. The yield can be increased by use of columns with larger inner diameters. Established sizes for semi-preparative scale are inner diameters from 8 to 21 mm ID and from 21 to 80 mm ID on the preparative scale.

Alternatively, for smaller inner diameters the yield can be increased by higher throughput of samples (higher number of separation runs).

 

  Finally, an individual economical decision between the maximisation of purity, yield and throughput is necessary for the introduction of a preparative HPLC method.
  Scale-up from analytical separation to preparative scale

Prior to development of a preparative separation method, ideally an analytical method with a sufficient separation of main and side components should be established. The selection of the analytical column with regard to an efficient phase (selectivity, pH and pressure stability, loadability, availability of suitable particle sizes) should be made with the prospect of a scale-up to the preparative scale. The chromatographic conditions should be selected such, that the sample mixture is separated with a good resolution, preferably under isocratic conditions. Gradients are not recommended for preparative separations, because they always imply a great effort and with it higher operating costs. Good resolution for the analytical chromatogram offers the advantage, that the load on the preparative column can be higher. Hence, the yield can be increased.


For scale-up to the preparative scale a column with the same length but with a larger inner diameter is selected. In reference to the scale-up formula the loading amount of the sample is proportional to the cross-section of the column.

  M: sample mass
L: column length
d: column diameter
a: analytical column
p: preparative column


The eluent flow has to be adapted to the larger inner diameter to achieve similar retention times in comparison with the analytical column.
Due to overload effects the amount of sample loadable on the column is considerably increased; a simple up-scaling would be uneconomical with respect to eluent and adsorbent needs. One distinguishes concentration, volume and mass overload.

   
For concentration overload the concentration of the sample substances is increased while the injection volume remains constant. With increasing overload a gradually growing peak deformation results, which approximates the shape of a triangle. A concentration overload can only be realised, if the sample substances have sufficient solubility in the sample solution. Often DMSO is used as injection solvent, which requires a high stability of the column.

For low solubility of sample substances an overload can only be achieved by continuously increasing the injection volume with the same sample concentration.
The peaks approach the shape of a rectangle with this volume overload.

The term mass overload is used, if the injected sample mass (calculated from injection volume and concentration) exceeds a certain value, at which the local concentration of the sample in the column is so large, that an equilibration is no longer possible. The mass overload gives unintentional changing of the retention volumes of the peak fronts and peak broadening with tailing.


To get as much pure product per time unit as possible, the column is almost always operated in an overloaded status. A combination of concentration and volume overload is most often used in practice. Volume overload rather results for diluted sample solutions while concentration overload is observed for concentrated samples. In many cases both effects exist and the peaks become trapezoidal. For strongly concentrated samples, additionally mass overload may appear, which for columns with low loadability results in premature tailing. In general, the target is approximation to the case of concentration overload, because a greater sample amount can be separated.

The displayed up-scaling table shows typical sample amounts in relation to RP phases. Whereas 0.02 – 2 mg sample can be loaded onto an analytical column with 4 mm ID, the amount increases to 0.6 – 60 mg for a preparative column with 21 mm ID with the linear scale-up factor of 27.6, and to 3 – 350 mg for 50 mm ID with factor 161.3.

   
 
Of course the stated maximal amounts always depend on the separation problem, the sample composition and the quality of the preparative column in relation to loadability, selectivity and base deactivation. The mentioned values can be under-run as well as considerably exceeded.

Demands on a modern column for an efficient and economical preparative HPLC

To achieve a yield as high as possibly by overload, the column should have a high loadability. The highly selective and base deactivated NUCLEODUR® C18 HTec – especially developed for preparative applications – shows a considerably higher loading capacity in comparison with conventional phases. The “conventional phase” with a very good characteristic for analytical separations, used in the presented example for the comparison, already shows no baseline separation for a medium load.
In contrast NUCLEODUR® C18 HTec still features a clear separation and guarantees a high yield and purity.





Besides an adequate selectivity, which allows optimal separations, good base deactivation is important. If the analytical separation already shows tailing effects due to poor deactivation, then the preparative column prematurely reaches the load limits under overload conditions. This means that maximal yield and purity of the product cannot be achieved.

A high lifetime of the column is another factor influencing the economy of a preparative HPLC method. This is particularly important, if a maximal yield requires a high sample throughput. Hence the column should feature a good lifetime, even for an application with extreme sample solvents, extreme pH values, increased temperatures or with high flow rates, which an efficient operation sometimes may demand.

   
 
NUCLEODUR® C18 HTec is a reliable and durable RP phase with exceptional stability, which realises highest lifetime, even under challenging conditions.
Like every preparative phase from MACHEREY-NAGEL it is offered in VarioPrep columns with axially adjustable end fittings. If the adsorbent bed should settle after some time by strong chemical and mechanical stress on the column material, the dead volume can be compensated with a readjustment of the endfittings. The performance of the column will be completely reconstituted and its lifetime will be considerably extended.
Furthermore the lifetime can be extended with the use of guard columns. The new VarioPrep guard column system is a cost-efficient cartridges system, which features easy handling without impairment of the separation efficiency of the main column.


Conclusion

For an efficient and economical operation of a preparative HPLC method, column dimensions and phase parameters have to be adapted to individual requirements concerning purity, yield and throughput. A high loadability, optimal selectivity and good base deactivation of the phase enhance optimisation of purity and yield. Moreover lifetime of the preparative column can be considerably extended with a high chemical and mechanical stability of the phase, with an adjustable column system and with the use of guard columns, thus reducing costs.


Contact to the author - Please click here!
 Print   
     
Special Offer

25% Discount on NUCLEODUR® C18 HTec HPLC Columns

Try our new NUCLEODUR® C18 HTec and receive 25% introductory discount
 

MACHEREY-NAGEL offers exclusively for our Chroma-News readers a special introductory discount of 25% on any order of NUCLEODUR® C18 HTec columns. Choose any analytical or preparative column and profit from this exclusive offer.

Due to innovative surface coating procedures NUCLEODUR® C18 HTec provides excellent analytical separation properties and is the first choice for up-scaling to preparative column dimensions.

Just contact MN by referring to promotion code HTec and we will arrange all necessity with your local MN distributor.

To take advantage of this SPECIAL OFFER – Please click here!

This offer is valid until 30th April 2010 and exclusive for listed Chroma-News recipients - not valid for distributors; limited to one column per customer – while stocks last!

 Print   
     
MN Application Highlight

TLC separation of saffron ingredients with the
new ALUGRAM®
Xtra SIL G

Whenever a rapid and economical analysis from complex matrix is carried out, thin layer chromatography is the method of choice. Especially for the examination of food, where coloured or stained ingredients must be identified, a precise colour recognition is important.

The TLC separation of saffron ingredients with the new ALUGRAM® Xtra SIL G with its optimised binder system, shows even with polar aqueous eluents the definite correlation of naphthol yellow S as main ingredient.

View FULL APPLICATION NOTE MN Appl. No. 403600 – Please click here!

You can find more than 3000 further application notes in the MN application database! 

  
  
     

MN Product Highlight


VarioPrep Guard Columns - Perfect Protection of your Prep-Column

 

MACHEREY-NAGEL offers a new VarioPrep Guard Column system which provides perfect protection of your preparative main columns.

The cartridges are filled with specially fractionised NUCLEODUR® and NUCLEOSIL® silica, suitable for all MACHEREY-NAGEL columns as well as for various prep-columns from other manufactures.

The robust cartridge system offers:

  • easy, quick and simple handling and cartridge exchange
  • minimal invasive – no negative impact on separation efficiency of main column
  • lowest increase of back pressure
  • cost efficient cartridge system

VarioPrep Guard Columns are available in different dimensions (length/ID in mm) matching the individual needs of your preparative main column:

  • VP  10/8  for prep columns with 8 – 10 mm ID    (requires holder REF 718251)
  • VP 20/16 for prep columns with 16 – 21 mm ID  (requires holder REF 718250)
  • VP 15/32 for prep columns with ≥ 32 mm ID       (requires holder REF 718253)


More INFORMATION about VarioPrep column hardware – Please click here!

 Print   
     
 
 MN Literature Recommendation

Reversed Phase HPLC Application Guide – A Technical Handbook

MACHEREY-NAGEL, the manufacturer of NUCLEODUR® and NUCLEOSIL®, and one of the leading companies for chromatography products, offers a special application guide for reversed phase HPLC.


The 160-page publication includes:

  • theoretical aspects of HPLC
  • demands on high performance silica
  • stationary phases in RP-chromatography
  • retention and selectivity
  • mobile phase
  • particle size, flow rate and separation efficiency
  • do’s and don’ts – troubleshooting
  • basic aspects of preparative HPLC



More than 150 up-to-date and relevant applications cover these 5 important application fields:

  • drugs
  • biological and natural compounds
  • food analysis
  • environmental analysis
  • organic compounds

This guide is based on more than 30 years of experience and proving MN´s expertise in HPLC. Take the chance, profit from this experience and improve your knowledge and HPLC skills!


Click here to find out more about our broad range of NUCLEODUR®
and NUCLEOSIL® HPLC columns

Order a HARDCOPY of the HPLC Application Guide – Please click here! 

You can find more than 3000 application notes on the MN application database!




 Print   
     

 

Contact MN

International contact

List of all international MN DISTRIBUTORS and subsidiaries - Please click here! 

They will be glad to help you with your questions concerning our chromatography products.

 




Please visit us at analytica 2010 in Munich (Germany), March 23rd to 26th 2010. Don´t miss the presentation of our latest products. We are happy to welcome you at our booth.

Hall A2 - Booth 330


Any feedback, ideas, questions? … please contact us directly at info@mn-net.com

  
     
 Print   
   
  
Copyright 2009 by MACHEREY-NAGEL GmbH & Co. KG   Imprint / Impressum Privacy StatementTerms Of Use