Login
GC Troubleshooting
We can help you.
Click on the respective observation (chromatogram) to display the possible causes
and their remedies.
no peaks
1. no Peaks
missing peaks
2. Missing or overlapping peaks, poor separation efficiency
too small peaks
3. Peaks too small, poor quantification, concentrations not reproducible
increasing retention
4. Increased or differing retention times/ low gas flow
decreasing retention
5. Decreased or differing retention times
declining baseline
6. Constantly declining baseline
rising baseline
7. Constantly rising baseline
bleeding
8. Increasing baseline, at high temperature bleeding or noise
plateaus
9. Plateaus at certain temperatures
interfering peaks
10. Regular interfering peaks
spikes, ghost peaks
11. Irregular interfering peaks, spikes, ghost peaks
strong noise, waves
12. Strong noise, waves
peaks on hill
13. Peaks on fronting-hill, peaks on tailing-hill
fronting, tailing
14+15. Fronting or Tailing
broad peaks
16. Broad peaks
cut tops
17. Cut tops of peaks, broad peaks
negative peaks
18. Negative peaks, negative signals
double peaks
19. Double peaks, doubled peak tops
poor resolution
20. Short lifetime, poor resolution, lack of separation efficiency
baseline shift
21. Baseline increases or decreases before or after a peak

1. Observation: no peaks

Possible causes Suggested remedy
1. error in the detector power supply / electronics. 1. check detector/electronic power supply and cables.
2. no FID flame.
2. check FID; reignite it.
3. syringe defective / clogged. 3. use a different syringe or clean it.
4. temperature too low for the analytes, oven heating faulty. 4. check temperature program, oven temperature.
5. detector / software / computer hardware failure. 5. check integrator, cables; restart computer.
6. no gas flow. 6. check gas tubes, valves, seals; test gas flow; shorten front of GC column, change injection septum.
7. column connection leaks. 7. use new ferrules.
8. broken GC column. 8. if breakage is at the beginning or at the end, remove the short piece; breakage in the middle can be mended with a glass connector; for multiple breakages: replace column.
top of page

2. Observation: Missing or overlapping peaks, poor separation efficiency

2. Missing or overlapping peaks, poor separation efficiency Possible causes Suggested remedy
1. syringe defective / clogged. 1. use a different syringe or clean it.
2. sample too diluted. 2. increase injection volume; concentrate sample.
3. sample concentration too high. 3. decrease injection volume; dilute sample.
4. column connection leaks, column not properly installed 4. check column installation; search for leaks; replace ferrules.
5. perforated injection septum. 5. replace septum.
6. injector temperature too low. 6. check temperature program; increase injector temperature.
7. sample decomposes in the injector. 7. check temperature program; reduce injector temperature; replace liner; check capillary ends.
8. column oven too hot. 8. check temperature program, oven temperature (external thermometer); decrease temperature.
9. incorrect flow rate. 9. measure flow, control and adjust it if necessary.
10. column absorbs or decomposes analytes.

10. check capillary ends; check intact deactivation using the test mixture; for poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects: a) use a column with thicker film
b) use phase with better deactivation
c) use column with special selectivity

top of page

 

3. Observation: Peaks too small, poor quantification, concentrations not reproducible

3. Peaks too small, poor quantification, concentrations not reproducible Possible causes Suggested remedy
1. dirty syringe. 1. use a different syringe or clean it.
2. concentration of sample too low. 2. increase injection volume; concentrate sample.
3. split too high. 3. reduce split.
4. sensitivity of detector too low. 4. inject standard in order to test detector sensitivity.
5. column connection leaks, column not properly installed. 5. check column installation; search for leaks; replace ferrules.
6. injector temperature too low. 6. check temperature program, increase injector temperature.
7. dirty ECD. 7. clean ECD.
8. FID, TCD gas flow too low. 8. correct flow according to manufacturers’ instructions.
9. sample decomposes. 9. check capillary ends; check intact deactivation using the test mixture; for poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects: a) use a column with thicker film
b) use phase with better deactivation
c) use column with special selectivity.
top of page

4. Observation: Increased or differing retention times / low gas flow

4. Increased or differing retention times/ low gas flow Possible causes Suggested remedy
1. speed of gas too low. 1. increase flow.
2. column connection leaks, column not properly installed. 2. check column installation; search for leaks; replace ferrules.
3. oven temperature too low or unstable. 3. check temperature program, oven temperature (external thermometer); if the analytes are stable, increase temperature.
4. strong decrease of gas pressure. 4. replace septum; for an instrument with pressure / temperature control, flow pressure must be higher than 15 psi above the demand at max. temperature of the program. 
5. tubes / capillaries / column constricted or blocked. 5. compare flow at column entrance and outlet with preset flow; check and / or clean gas tubes; in case of pressure build-up cut and remove one turn
(20 cm) from the column or replace column.
top of page

5. Observation: Decreased or differing retention times

5. Decreased of differing retention times Possible causes Suggested remedy
1. speed of gas too high. 1. compare flow at column entrance and outlet with preset flow; check gas tubes and pressure gauge; control parameter settings; or replace column.
2. oven temperature too high. 2. check temperature program, oven temperature (external thermometer); decrease temperature.
3. column length too short. 3. replace column.
4. film thickness in column too low. 4. replace column.
top of page

6. Observation: Constantly declining baseline

6. Constantly declining baseline Possible causes Suggested remedy
1. gas flow changes with temperature gradient. 1. check gas content in gas cylinder; pressure must be a few bar above the required pressure at max. temperature; otherwise exchange gas cylinder.
2. contaminated gas/ poor gas quality
(at constant inlet pressure).
2. check gas supply.
3. column not properly installed. 3. check column installation (FID).
top of page

7. Observation: Constantly rising baseline

Possible causes Suggested remedy
1. leak at column entrance or injection septum. 1. check column installation; search for leaks; replace ferrules.
2. injector contaminated. 2. make a run at lower injector temperature; if the baseline improves, replace liner, use low bleed or high temperature septa.
3. column contaminated. 3. cut two turns from column entrance; rinse column with solvent (only chemically bonded phases); otherwise replace column or use guard column.
4. detector contaminated. 4. clean detector.
5. increase of temperature too fast. 5. decrease temperature gradient and end temperature.
6. poor gas quality. 6. use gas grades recommended for GC; for longer supply lines from gas source to GC use gas purification cartridges directly connected to the GC.
top of page

8. Observation: Increasing baseline, at high temperature bleeding or noise

8. Increasing baseline, at high temperature bleeding or noise Possible causes Suggested remedy
1. decomposition of the stationary phase. 1. check for leaks; matrix check for compatibility with the column.
2. column contaminated. 2. cut two turns from column entrance; rinse column with solvent (only chemically bonded phases); otherwise replace column or use guard column.
3. increase of temperature too fast / end temperature too high. 3. decrease temperature gradient and end temperature.
4. column not properly conditioned. 4. condition column according to manufacturers’ instructions (while column is not connected to the detector).
5. detector contaminated. 5. clean detector according to manufacturers’ instructions.
6. poor gas quality. 6. use gas grades recommended for GC; for longer supply lines from gas source to GC use gas purification cartridges directly connected to the GC.
top of page

 

9. Observation: Plateaus at certain temperatures

9. Plateaus at certain temperatures Possible causes Suggested remedy
1. steps in temperature program too drastic. 1. avoid very short and strong heating periods.
top of page

 

10. Observation: Regular interfering peaks

10. Regular interfering peaks Possible causes Suggested remedy
1. poor gas quality. 1. use gas grades recommended for GC; for longer supply lines from gas source to GC use gas purification cartridges directly connected to the GC.
2. FID: dust or contaminants in the detector. 2. clean detector; if particles are visible in the column or column ends are not cut precisely (frayed edges), cut two turns from the column entrance.
3. electronic defect, damaged cable or detector. 3. replace cable, contact your GC manufacturer.
4. bleeding of silicon septa. 4. replace injection septum, use low bleed or high temperature septa.
top of page

 

11. Observation: Irregular interfering peaks, spikes, ghost peaks

11. Irregular interfering peaks, spikes, ghost peaks Possible causes Suggested remedy
1. contamination from vials / septa or sample preparation. 1. control SPE and / or autosampler vials; use low bleed or high temperature septa.
2. derivatization not quantitative. 2. check derivatization protocol; use more reactive derivatization reagents.
3. dirty syringe. 3. use a different syringe or clean it.
4. sample decomposes. 4. check temperature program, oven temperature (external thermometer); if analytes are not temperature-stable, reduce injector temperature; replace liner.
5. column absorbs or decomposes analytes.

5. check capillary ends; check intact deactivation using the test mixture; for poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects:
a) use a column with thicker film
b) use phase with better deactivation
c) use column with special selectivity.

6. sample volume too high, double injection. 6. reduce sample volume or add a blank run after a high volume injection.
7. poor gas quality. 7. use gas grades recommended for GC; for longer supply lines from gas source to GC use gas purification cartridges directly connected to the GC.
top of page

 

12. Observation: Strong noise, waves

12. Strong noise, waves Possible causes Suggested remedy
1. leak at column entrance or injection septum. 1. check column installation; search for leaks; replace ferrules.
2. bleeding of septum / injector contaminated. 2. make a run with lower injector temperature; if the baseline improves, replace liner, use low bleed or high temperature septa.
3. septum particles in column entrance. 3. cut 1 turn from column entrance; replace injection septum.
4. column contaminated. 4. cut two turns from column entrance; rinse column with solvent (only chemically bonded phases); otherwise replace column or use guard column.
5. column not properly conditioned.

5. condition column according to manufacturers’ instructions (while column is not connected to the detector).

6. hardware defect. 6. check temperature program, oven temperature (external thermometer); contact your GC manufacturer.
7. detector contaminated (electronics). 7. clean detector according to manufacturers’ instructions; check electronics.
8. increase of temperature too fast. 8. decrease temperature gradient and end temperature.
9. poor gas quality. 9. use gas grades recommended for GC; for longer supply lines from gas source to GC use gas purification cartridges directly connected to the GC.
top of page

13. Observation: Small peaks on fronting or tailing of bigger peaks

13. Peaks on fronting-hill, peaks on tailing-hill Possible causes Suggested remedy
1. column not properly installed. 1. check capillary ends; check tight and correct fit in injector and detector.
2. temperature of injection too low. 2. check injector temperature; if analytes are stable, increase temperature.
3. solvent not compatible with GC phase. 3. change solvent.
4. splitter defect. 4. measure flow and adjust splitter.
5. poorly deactivated column, film thickness too low. 5. check capillary ends; check intact deactivation using the test mixture; for poor results shorten both column ends by about 10 cm; or replace column.
top of page

 

14. Observation: Fronting (strong broadening in the ascending part)

Fronting14. Fronting (strong widening in the ascending part) Possible causes Suggested remedy
1. column overload. 1. decrease injection volume; dilute sample.
2. sample vaporizes too slowly, not evenly or condenses. 2. increase injector temperature (consider max. temperature limits of the column).
3. analytes coelute. 3. change temperature program or use column with different selectivity.
4. sample decomposes. 4. check temperature program, oven temperature (external thermometer); if analytes are not temperature-stable, reduce injector temperature; replace liner.
5. column absorbs or decomposes analytes.

5. check capillary ends; check intact deactivation using the test mixture; for poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects: a) use a column with thicker film
b) use phase with better deactivation
c) use column with special selectivity.

15. Observation: Tailing (strong broadening in the descending part)

Tailing15. Tailing  (strong widening in the descending part) Possible causes Suggested remedy
1. sample vaporizes too slowly, not evenly or condenses. 1. increase injector temperature (consider max. temperature limits of the column).
2. high-boiling analytes. 2. derivatize polar, basic or high-boiling compounds.

3. system leaks.

3. check column installation; search for leaks; replace ferrules.
4. analytes coelute. 4. change temperature program or use column with different selectivity.
5. sample decomposes.

5. check temperature program, oven temperature (external thermometer); if analytes are not temperature-stable, reduce injector temperature; replace liner by a deactivated one.

6. column absorbs or decomposes analytes. 6. check capillary ends; check intact deactivation using the test mixture; for poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects: a) use a column with thicker film
b) use phase with better deactivation
c) use column with special selectivity.
7. split rate too low. 7. increase split rate.
8. analytes always tending to tail. 8. no chance for symmetric peaks.
9. column overload. 9. decrease injection volume; dilute sample.

 

top of page

 

16. Observation: Broad peaks

16. Broad peaks Possible causes Suggested remedy
1. poor focussing. 1. decrease start temperature of the program.
2. flow too high or too low. 2. measure flow, control and adjust it if necessary.

3. split rate too low.

3. increase split rate.
4. column overloaded. 4. decrease injection volume, dilute sample or increase split flow.
top of page

 

17. Observation: Cut tops of peaks, broad peaks

17. Cut tops of peaks, broad peaks Possible causes Suggested remedy
1. detector overloaded. 1. decrease injection volume; dilute sample; increase the split flow.
2. column overloaded. 2. decrease injection volume; increase split flow.

3. zero point is outside the display.

3. change scale.
top of page

 

18. Observation: Negative peaks, negative signals

18. Negative peaks, negative signals Possible causes Suggested remedy
1. polarity of integrator is inverted. 1. invert polarity at the instrument.
2. column overload. 2. decrease injection volume; dilute sample / increase split rate.
3. pressure fluctuations. 3. check gas tubes, valves, seals; test gas flow; change injection septum; contact hardware manufacturer.
4. detector contaminated. 4. clean detector as specified by the manufacturer.
5. electronic artefacts. 5. check detector, A/D converter.
top of page

 

19. Observation: Double peaks, doubled peak tops

19. Double peaks, doubled peak tops Possible causes Suggested remedy
1. solvent and column not compatible. 1. change solvent or use guard column.
2. solvent mixtures with large differences in boiling point and polarity. 2. use just one solvent.

3. sample decomposes.

3. check temperature program, oven temperature (external thermometer); if analytes are not temperature-stable, reduce injector temperature; replace liner by a deactivated one.
4. analytes coelute. 4. modify temperature program or use longer column; possibly change column polarity.
5. detector overload.

5. inject less; control make-up flow.

top of page

 

20. Observation: Short lifetime, poor resolution, lack of separation efficiency

20. Short lifetime, poor resolution, lack of separation efficiency Possible causes Suggested remedy
1. impurities on the column. 1. cut two turns from column entrance; rinse column with solvent (only chemically bonded phases); otherwise replace column or use guard column.
2. contamination from vials / septa or sample preparation. 2. check SPE and / or autosampler vials; use low bleed or high temperature septa.

3. polymerization on the column.

3. use guard column (at least 10 m).
4. separation efficiency decreases for repeated injections, improves after reconditioning. 4. use guard column; reduce injection volume; modify temperature program; for repeated injections increase end temperature (if possible) and use longer temperature program. 
5. temperature too high / temperature increase too fast.

5. decrease oven temperature and / or temperature gradient (should not be higher than 25 °C/min).

6. cooling too fast. 6. do not open oven door at high temperatures.
7. temperature too low / condensation. 7. increase injector temperature and/or start temperature.
8. poor deactivation. 8. check capillary ends; check intact deactivation using the test mixture; for poor results shorten both column ends by about 10 cm; or replace column; if column test does not show any defects:
a) use a column with thicker film
b) use phase with better deactivation
c) use column with special selectivity.
9. oxygen / air in the system. 9. use oxygen absorber or gas grade with less oxygen.
10. water content too high. 10. reduce water content.
11. head-space analysis: permanent air injections. 11. displace oxygen from vials with an inert gas.
top of page

 

21. Observation: Baseline increases or decreases before or after a peak

21. Baseline increases or decreases before or after a peak Possible causes Suggested remedy
1. injection volume too high. 1. decrease injection volume; dilute sample; clean injection system.
2. column bleeding due to poor conditioning. 2. condition column according to manufacturers’ instructions (while column is not connected to the detector).

3. pressure fluctuations.

3. check gas tubes, valves, seals; test gas flow; change injection septum; contact hardware manufacturer.
4. injector temperature too low. 4. check injector temperature; if the analytes are stable, increase temperature.
5. injection septum perforated.

5. replace septum.

6. wrong TCD gas flow. 6. adjust flow according to manufacturers’ instructions.
top of page

Copyright 2013 by MACHEREY-NAGEL GmbH & Co. KG   Imprint / Impressum Privacy StatementTerms Of Use